Teledyne API Photometric O₃ Calibrator (Model T703)

Teledyne API Photometric O₃ Calibrator (Model T703)

A microprocessor-controlled ozone calibrator for precision gas analysers, that can deliver concentrations from 50 ppb to 5 ppm at flows from 1 LPM to 5 LPM

How it works

The Teledyne API Photometric O₃ Calibrator (Model T703) is a microprocessor-controlled ozone calibrator for precision gas analysers, that can deliver concentrations from 50 ppb to 5 ppm at flows from 1 LPM to 5 LPM

The Model T703 has two major components; the ozone generating device and the photometer. The ozone generator produces ozone by passing zero air over a UV lamp. Flow rate and lamp intensity are adjusted to deliver a specified ozone concentration by volume.

The temperate and pressure compensated photometer, identical to that of the Model T400 ozone analyser, provides feedback control of the UV lamp intensity, assuring stable ozone output. A nearly unlimited number of calibration sequences may be programmed into the non-volatile memory of the T703, covering time periods up to a year.

Servicing

Servicing information

Enquire about this product

Please enable JavaScript in your browser to complete this form.
Address