How it works
The Teledyne API MCERTS N500 CAPS NOx Analyser (Nitrogen Oxides) uses superior Cavity Attenuated Phase Shift (CAPS) Spectroscopy to measure True NO₂, NOx, and NO gases.
The instrument combines direct NO₂ measurements with highly efficient gas phase titration (GPT) to convert and measure the NO gas component.
Servicing
Data sheets and specifications are subject to change, and Enviro Technology cannot be held responsible for inaccuracies in manufacturer-issued data sheets. To confirm specifications and obtain the latest issue of the data sheet, please contact us.
Datasheet
Specifications
- Measurement Units ppb, ppm, μg/m³, mg/m³ (user-selectable)
- Response Time < 60 seconds to 95%
- Ranges Min: 0 – 5 ppb full scale
- Max: 0 – 1,000 ppb full scale
- Sample Flow Rate 1000 cc/min ±10%
- Zero Noise < 0.05 ppb (RMS)
- Span Noise < 0.2% of reading + 50 ppt (RMS)
- Lower Detectable Limit < 0.1 ppb
- Precision 0.5% of reading above 5 ppb
- Linearity 1% of full scale
- Zero Drift < 0.2 ppb / 24 hours
- Span Drift < 0.5% of reading / 24 hours
- Included I/O 1 x Ethernet (TCP/IP)
- 1 x RS232
- 2 x front panel USB device ports